Model | BK-200 |
Overall Performance | Throughput | 200 Tests/hour |
Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry |
Certificates | CE, FDA, ISO9001, ISO14001, ISO13485 |
Sample & Reagent System | Sample Positions | 37 sample positions |
Reagent Position | 28 reagent positions |
Sample Volume | 2-70µl |
Reagent Volume | 20-350µl |
Probe | Teflon coating, with anti-collision function, liquid level detection function |
Probe Washing | Automatic washing interior and exterior |
Reagent Cooling | Refrigerated tray with independent switch |
Reaction System | Temperature Control | 37+0.2℃, real-time monitoring |
Cuvettes | 48 reusable cuvettes, optical length 6mm |
Mixer Probe | Independent stirring |
Washing | Automatic cuvettes washing |
STAT function | YES |
Optical System | Light Source | 6V/10W halogen lamp |
Spectrophotometry | Post-spectral spectrophotometry |
Wavelength | 340,405,450,480,505,546,570,600,,630,700,750,800nm |
Absorbance | 0-3.0Abs |
Calibration & QC | Calibration | Linear:K factor, 1-point ,2-point and multipoint linear Non-Linear: Spline,Polygon, Index, Ogarithm, Logit-4P, Logit-5P |
Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot(2D) |
Data Management | Software | Windows 7/8/10, 32 or 64bit |
LIS System | Bi-direction, support HL7 protocol |
Interface | LAN port access |
Printer | External, multiple reporting mode available |
Working Conditions | Power Supply | AC220V±10%, 50/60Hz, 110V±10%, 60Hz, 300W |
Temperature | 15~30℃ |
Water Consumption | Deionized water: <5L/H |
Humidity | 40-85% |
Size & Weight | External Size (W*D*H) | 625*425*460mm |
Net Weight | 36kg |
Package Size (W*D*H) | 750*560*900mm |
Gross Weight | 55kg |